VA-SOP-LC-006 · Standard operating procedure

Reprocessing data files with a new method in LCsolution Postrun Batch

SOP with screenshots: reprocess all injections with a corrected integration method using the LCsolution Postrun Batch Table, without re-injecting.

SOP No.VA-SOP-LC-006
Revision00
Effective date30 October 2026
Review date30 October 2027
SoftwareShimadzu LCsolution 1.25 (Postrun Analysis), LC-2010 series HPLC

1Purpose

To apply a revised data-analysis method to existing data files in one step using the Postrun Batch Table.

2Scope

Applies to LCsolution 1.2x Postrun Batch Processing.

3Responsibility

  • Analyst — performs the procedure and records results.
  • QC reviewer — verifies the output against the acceptance criteria.
  • Lab in-charge / system administrator — controls method and report-format files and user rights.

5Prerequisites and materials

  • Revised method file (.lcm) saved.
  • Data files to reprocess, preferably copies.

6Safety and precautions

  • Work on a copy of data and method files until the change is approved.
  • Never delete or overwrite raw data (.lcd) to obtain a passing result; every change must be traceable.
  • Follow the laboratory's data-integrity and audit-trail SOP when LCsolution security is enabled.

7Procedure

  1. 7.1

    Set the method in the Batch Table

    Postrun → Batch Processing: put the new method (e.g. VA_Caffeine.lcm) in the Method File column; Edit → Fill Down copies it to every row.

    Step 7.1: Set the method in the Batch Table — Shimadzu LCsolution screenshot
    Fig. 7.1 — Set the method in the Batch Table
  2. 7.2

    Start the batch

    Click Batch Start. If a data file is open, the "Close Read File" prompt appears — click OK.

    Step 7.2: Start the batch — Shimadzu LCsolution screenshot
    Fig. 7.2 — Start the batch
  3. 7.3

    Check completion

    All six data files are reprocessed in seconds.

    Step 7.3: Check completion — Shimadzu LCsolution screenshot
    Fig. 7.3 — Check completion
  4. 7.4

    Open one injection

    Double-click any file, e.g. Caffeine_04.

    Step 7.4: Open one injection — Shimadzu LCsolution screenshot
    Fig. 7.4 — Open one injection
  5. 7.5

    Verify the new integration

    Impurity separated, main peak Area % 82.7.

    Step 7.5: Verify the new integration — Shimadzu LCsolution screenshot
    Fig. 7.5 — Verify the new integration
  6. 7.6

    Good practice

    Work on a copy; File → Rollback to Original Data gives a clean start.

8Acceptance criteria and interpretation

  • Every row of the batch shows the new method and finishes without error.
  • Opened data files show the new integration (e.g. impurity separated, Area % 82.7).

9Troubleshooting and common errors

SymptomCause and fix
"Close Read File" messageA data file is open in another window. Click OK; LCsolution closes it first.
Need the original result backFile → Rollback to Original Data restores the acquisition-time processing.

10References

  • Shimadzu LCsolution Ver. 1.2x — Operation and Help documentation.
  • USP <621> Chromatography.
  • ICH Q2(R2) Validation of Analytical Procedures.

11Revision history

Rev.DateChange
0030 October 2026First issue, written from the October 2026 screen recording.

Frequently asked

How do I reprocess HPLC data in LCsolution without injecting again?

Put the new method in the Method File column of the Postrun Batch Table (Edit → Fill Down) and click Batch Start.

Video version: LCsolution — Changed your integration method? Reprocess all six injections (October 2026) — on our Instagram and YouTube.

Related SOPs

Vision Analytical is an independent service provider, not affiliated with Shimadzu Corporation or Waters Corporation. LCsolution and Empower are trademarks of their respective owners. Screenshots are from our own demonstration systems with demonstration data. Last updated 30 October 2026.

Contact

Talk to an engineer, not a call centre

Quotes, spare-part availability and breakdown support — reach us whichever way suits you.

+91 9819424897EmailGet a quote