VA-SOP-LC-006 · Standard operating procedure
Reprocessing data files with a new method in LCsolution Postrun Batch
SOP with screenshots: reprocess all injections with a corrected integration method using the LCsolution Postrun Batch Table, without re-injecting.
| SOP No. | VA-SOP-LC-006 |
|---|---|
| Revision | 00 |
| Effective date | 30 October 2026 |
| Review date | 30 October 2027 |
| Software | Shimadzu LCsolution 1.25 (Postrun Analysis), LC-2010 series HPLC |
1Purpose
To apply a revised data-analysis method to existing data files in one step using the Postrun Batch Table.
2Scope
Applies to LCsolution 1.2x Postrun Batch Processing.
3Responsibility
- Analyst — performs the procedure and records results.
- QC reviewer — verifies the output against the acceptance criteria.
- Lab in-charge / system administrator — controls method and report-format files and user rights.
5Prerequisites and materials
- Revised method file (.lcm) saved.
- Data files to reprocess, preferably copies.
6Safety and precautions
- Work on a copy of data and method files until the change is approved.
- Never delete or overwrite raw data (.lcd) to obtain a passing result; every change must be traceable.
- Follow the laboratory's data-integrity and audit-trail SOP when LCsolution security is enabled.
7Procedure
-
7.1
Set the method in the Batch Table
Postrun → Batch Processing: put the new method (e.g. VA_Caffeine.lcm) in the Method File column; Edit → Fill Down copies it to every row.

Fig. 7.1 — Set the method in the Batch Table -
7.2
Start the batch
Click Batch Start. If a data file is open, the "Close Read File" prompt appears — click OK.

Fig. 7.2 — Start the batch -
7.3
Check completion
All six data files are reprocessed in seconds.

Fig. 7.3 — Check completion -
7.4
Open one injection
Double-click any file, e.g. Caffeine_04.

Fig. 7.4 — Open one injection -
7.5
Verify the new integration
Impurity separated, main peak Area % 82.7.

Fig. 7.5 — Verify the new integration -
7.6
Good practice
Work on a copy; File → Rollback to Original Data gives a clean start.
8Acceptance criteria and interpretation
- Every row of the batch shows the new method and finishes without error.
- Opened data files show the new integration (e.g. impurity separated, Area % 82.7).
9Troubleshooting and common errors
| Symptom | Cause and fix |
|---|---|
| "Close Read File" message | A data file is open in another window. Click OK; LCsolution closes it first. |
| Need the original result back | File → Rollback to Original Data restores the acquisition-time processing. |
10References
- Shimadzu LCsolution Ver. 1.2x — Operation and Help documentation.
- USP <621> Chromatography.
- ICH Q2(R2) Validation of Analytical Procedures.
11Revision history
| Rev. | Date | Change |
|---|---|---|
| 00 | 30 October 2026 | First issue, written from the October 2026 screen recording. |
Frequently asked
How do I reprocess HPLC data in LCsolution without injecting again?
Put the new method in the Method File column of the Postrun Batch Table (Edit → Fill Down) and click Batch Start.