Get a quote

Shimadzu HPLC Service and Repair

Service, repair, spares and AMC for Shimadzu LC-2010, LC-20 Prominence and Nexera-i HPLC systems in Mumbai, Thane, Pune and Ahmedabad.

Key facts

Based in
Ambernath, Thane, Maharashtra · since 2022
Service area
Maharashtra & Gujarat
Mumbai · Thane · Ambernath · Pune · Navi Mumbai · Ahmedabad
Instruments
HPLC, GC, LC-MS and UV-Vis
Brands
Shimadzu, Waters, Agilent Technologies, Thermo Scientific, Merck, PerkinElmer
Services
Installation · Preventive Maintenance · Breakdown Service · AMC · CMC · Calibration · IQ · OQ · PQ · Validation · Software Installation · Training
Reach us
+91 9819424897 (phone) · +91 9819424897 (WhatsApp) · visionanalytical96@gmail.com · customer support info@visionanalytical.co.in
Monday to Saturday, 9:00 am to 6:30 pm IST, closed on Sunday

Shimadzu liquid chromatographs we service

We service Shimadzu HPLC systems used in pharmaceutical QC and research laboratories, including the LC-2010 and LC-2010C HT integrated systems, LC-20 Prominence modular systems and the Nexera-i LC-2030C. Work covers installation, preventive maintenance, breakdown repair, calibration, IQ, OQ and PQ qualification, and annual maintenance contracts. Shimadzu GC-2010 and GC-2014 gas chromatographs and the UV-1800 spectrophotometer are serviced by the same team, which helps laboratories that run several Shimadzu instruments.

Faults we see most often

On the LC-2010 family the common faults are pump pressure that fluctuates or falls because of worn plunger seals or a sticking check valve, autosampler carryover from a worn rotor seal or needle seat, and a deuterium lamp near the end of its life that raises baseline noise. Error messages on the front panel are often the fastest clue. Our troubleshooting guides list the checks we make in order, and our case study on an LC-2010 pump cooling fan error shows how a fault is traced to one part before anything is replaced.

Spares and repair

Wear parts for Shimadzu HPLC pumps, injectors and detectors, such as plunger seals, check valves, rotor seals, lamps and fittings, can be supplied with the repair. Older systems whose parts are hard to source are assessed honestly, and if a repair is not worth the cost we say so and explain the replacement options, including refurbished units. To raise a request, send the model and serial number of each module, the LabSolutions or software version, a photo of any error message on the front panel and a short note on when the fault started and what changed before it. That is usually enough to bring the right parts on the first visit.

Independent support across Maharashtra and Gujarat

Vision Analytical is an independent service company based in Ambernath, Thane district, and is not affiliated with or authorised by Shimadzu Corporation. Shimadzu, Prominence and Nexera are trademarks of Shimadzu Corporation and are used here only to describe the instruments we support. Engineers cover Mumbai, Thane, Navi Mumbai, Ambernath, Pune and Ahmedabad.

Which Shimadzu HPLC models does Vision Analytical service?

Vision Analytical services Shimadzu LC-2010 and LC-2010C HT integrated HPLC systems, LC-20 Prominence modular systems and the Nexera-i LC-2030C, along with Shimadzu GC-2010 and GC-2014 gas chromatographs and the UV-1800 spectrophotometer. Work includes installation, preventive maintenance, breakdown repair, spares, AMC and IQ, OQ and PQ qualification across Maharashtra and Gujarat.

Why does the pressure fluctuate on a Shimadzu LC-2010 pump?

Fluctuating pressure on an LC-2010 pump usually comes from air in the solvent line, a sticking or dirty check valve, or worn plunger seals. Purge the pump and check the degasser first, then clean or replace the check valves, and replace the seals if the pressure ripple stays. A leak at the pump head is a sign the seals are worn.

Contact

Talk to an engineer, not a call centre

Quotes, spare-part availability and breakdown support — reach us whichever way suits you.

+91 9819424897EmailGet a quote