VA-SOP-SRV-001 · Standard operating procedure

Preparing and issuing an instrument service report (HPLC / GC / UV-Vis)

How we prepare an HPLC, GC or UV-Vis service report — modules and serial numbers, fault, action, parts, status, sign-off — and what labs should check.

SOP No.VA-SOP-SRV-001
Revision00
Effective date30 October 2026
Review date30 October 2027
SoftwareVision Analytical service report form (paper at site, recorded in the office system; PDF copy issued)

1Purpose

To ensure that every service, breakdown, calibration, validation or AMC visit is recorded on one standard service report that the customer can file as evidence for audits (GMP / GLP / ISO 17025).

2Scope

Applies to all field visits by Vision Analytical engineers on HPLC, GC, UV-Vis and related laboratory instruments — new installation, calibration, demonstration, validation (IQ/OQ/PQ), preventive maintenance and repairs — under warranty, AMC, courtesy or paid visit.

3Responsibility

  • Service engineer — fills the report at site, explains the work, takes the customer's signature and stamp.
  • Customer representative — verifies the entries, writes comments, signs and stamps.
  • Service office — records the report the same day, issues the PDF copy and tracks open actions and next due dates.

4Definitions and abbreviations

AMCAnnual Maintenance Contract — scheduled preventive maintenance and breakdown support.
PMPreventive maintenance.
IQ / OQ / PQInstallation, Operational and Performance Qualification.
System No.Instrument or module serial number as printed on the instrument nameplate.

5Prerequisites and materials

  • Service call registered with a report number (format VA-SR-…).
  • Instrument make, model and serial numbers of every module worked on.
  • Calibration standards / tools with valid certificates when calibration or qualification is performed.

6Safety and precautions

  • Follow the customer's site safety and gowning rules; switch off and lock out power before opening modules.
  • Handle solvents, lamps (UV) and gas lines as per the customer's safety instructions.
  • Record only facts observed or done. Do not pre-fill results or sign on behalf of the customer.

7Procedure

    A. At site — header and instrument

  1. 7.1

    Report number and date

    Write the report number (from the registered call) and the visit date. One report per visit; continuation sheets carry the same number.

    Step 7.1: Report number and date — Service documentation screenshot
    Fig. 7.1 — Report number and date
  2. 7.2

    Customer details

    Company name and address, contact person, designation and department, telephone and weekly off — as the customer wants them on their records.

    Step 7.2: Customer details — Service documentation screenshot
    Fig. 7.2 — Customer details
  3. 7.3

    Time in / out and every module serviced

    One line per module — time in, time out, system configuration (pump, autosampler, detector, oven…), model and serial number from the nameplate. Serial numbers make the report traceable in your equipment log.

    Step 7.3: Time in / out and every module serviced — Service documentation screenshot
    Fig. 7.3 — Time in / out and every module serviced
  4. 7.4

    Status, type of call and contract

    Tick instrument status (OK / Not OK at the end of the visit), the type of call (installation, calibration, demonstration, validation, maintenance, repairs) and the contract (warranty, AMC, courtesy, paid visit).

    Step 7.4: Status, type of call and contract — Service documentation screenshot
    Fig. 7.4 — Status, type of call and contract
  5. B. At site — what was found and done

  6. 7.5

    Fault reported, observation and action taken

    Fault reported: the customer's complaint in their words (error code, symptom). Observation and action taken: what was measured (pressure, lamp energy, leak test, noise/drift values), the root cause, and the corrective action.

    Step 7.5: Fault reported, observation and action taken — Service documentation screenshot
    Fig. 7.5 — Fault reported, observation and action taken
  7. 7.6

    Parts replaced or required

    Part description, part number and quantity — replaced now, or required (quotation follows).

    Step 7.6: Parts replaced or required — Service documentation screenshot
    Fig. 7.6 — Parts replaced or required
  8. 7.7

    Customer comments and signatures

    The customer writes comments, signs with name, date and stamp; the engineer signs. Leave a copy with the customer.

    Step 7.7: Customer comments and signatures — Service documentation screenshot
    Fig. 7.7 — Customer comments and signatures
  9. C. Office — record and issue

  10. 7.8

    Issue the PDF and follow up

    The PDF copy goes to the customer by e-mail within one working day. Open points (parts required, re-visit) are tracked until closed; the next calibration / PM due date is noted for the AMC schedule.

8Acceptance criteria and interpretation

  • Every module worked on is listed with model and serial number.
  • Fault, observation/action and parts sections are filled in plain, factual language; measured values are written with units.
  • Status is ticked; customer and engineer signatures with date and stamp are present.
  • PDF copy issued within one working day and matches the signed paper copy.

9Troubleshooting and common errors

SymptomCause and fix
Customer not available to signNote the name of the person who witnessed the work; the signed copy is collected on the next visit and the PDF is marked 'awaiting customer signature'.
Work spans more than one dayOne report per visit day, each with its own time in/out; reference the previous report number.
Part not available at siteStatus 'Not OK', part listed under 'required'; quotation and re-visit date recorded.

10References

  • EU GMP Annex 15 / WHO TRS guidance on qualification and validation — records of maintenance and calibration.
  • ISO/IEC 17025:2017 clause 6.4 — equipment records.
  • USP <1058> Analytical Instrument Qualification.

11Revision history

Rev.DateChange
0030 October 2026First issue.

12Annexures

Annexure 1 — Vision Analytical service report form (blank).
Annexure 1 — Vision Analytical service report form (blank).

Frequently asked

What should an HPLC service report contain?

Report number and date, customer details, every module with model and serial number, call and contract type, fault reported, observation and action taken with measured values, parts replaced or required, instrument status, and customer and engineer signatures.

Is a service report enough for a GMP audit?

It is the primary maintenance record. Auditors also expect calibration certificates of the standards used and, after repairs that affect performance, an OQ/PQ or performance check.

Vision Analytical is an independent service provider, not affiliated with Shimadzu Corporation or Waters Corporation. LCsolution and Empower are trademarks of their respective owners. Screenshots are from our own demonstration systems with demonstration data. Last updated 30 October 2026.