Analytical flow cell for an HPLC UV or PDA detector. The cell carries the column eluate through the light path between two quartz windows; absorbance rises in direct proportion to the path length, which is why a standard analytical cell is 10 mm and why changing cell type changes apparent sensitivity even though nothing about the method has changed.
An analytical cell keeps its illuminated volume small - single-digit microlitres on most designs - so that the peak leaving the column is not broadened again inside the detector. Semi-preparative and high-sensitivity cells trade that volume against path length deliberately.
What goes wrong
- Deposits or air in the cell give exactly the same baseline noise as a dying lamp; check the cell first, because it costs nothing.
- High-pH mobile phases etch quartz windows, and an etched window shows as permanently low energy.
- A leaking cell leaves a salt crust at the fittings and drifts with room temperature.
- Over-tightened fittings crack windows and distort the flow path.
Cleaning is done by pumping the manufacturer's recommended solvent sequence through the cell - commonly isopropanol, or dilute nitric acid where the manual allows it - never by dismantling and wiping the windows.
- Serving laboratories since 2022
- Refurbished Instruments · Spare Parts Store · Service & AMC from the same team
- GST-registered supplier · GSTIN 27ABIPZ1312R1Z0
Specifications
| Cell type | Analytical flow cell |
|---|---|
| Typical path length | 10 mm |
| Illuminated volume | Single-digit microlitres (design dependent) |
| Windows | Quartz |
| Detectors | HPLC UV and PDA |
| Failure signs | Baseline noise, low energy, drift, salt crust at fittings |
| Listed for | Agilent, Shimadzu, Thermo Scientific, Waters |