Replacement check valve cartridge for the inlet or outlet position of an HPLC pump head. A check valve is a passive one-way valve - typically a ruby ball seating on a sapphire seat - that is opened and closed by the flow itself on every piston stroke, which is why it is one of the highest-cycle parts in the instrument.
Almost every check valve failure is caused by something sitting between the ball and the seat: a buffer crystal, a particle from an unfiltered mobile phase, or an air bubble that will not clear.
Symptoms of a failing check valve
- Pressure fluctuating in time with the pump stroke instead of holding steady.
- The pump losing prime, or refusing to re-prime after a solvent change.
- Flow that falls away while the set flow rate stays the same.
- Pressure that will not hold when the pump is stopped.
A cartridge that has only picked up a salt crystal can often be recovered by sonicating it in 50:50 methanol and water; one that fails again after cleaning has a worn seat and should be replaced. Inlet and outlet cartridges are not always interchangeable - check the flow direction marking before fitting.
- Serving laboratories since 2022
- Refurbished Instruments · Spare Parts Store · Service & AMC from the same team
- GST-registered supplier · GSTIN 27ABIPZ1312R1Z0
Specifications
| Type | Passive one-way cartridge valve |
|---|---|
| Typical construction | Ruby ball on sapphire seat |
| Positions | Pump head inlet or outlet |
| Main failure cause | Buffer crystals, particulates, trapped air |
| First aid | Sonicate in 50:50 methanol / water |
| Fit note | Observe flow direction marking |
| Listed for | Agilent, Shimadzu, Thermo Scientific, Waters |