Replacement injection syringe for HPLC and GC autosamplers, 100 µL nominal volume. The barrel is glass and the plunger is stainless steel, so the wearing parts are the plunger seal and the needle tip rather than the body.

A syringe ages by injection count and sample matrix, not by calendar date. As the plunger seal wears, the drawn volume stops being repeatable and the first symptom in the data is falling peak-area precision, not an error message.

What a worn syringe looks like

  • Peak-area %RSD rising across a sequence that used to pass.
  • Air bubbles drawn on aspiration, giving short or missed injections.
  • Carryover into the following injection from a scored needle tip.
  • Sticking plunger after sitting in a buffer that was left to dry.

Needle length, gauge and tip style (blunt, bevelled or conical) differ between autosampler designs even when the volume is the same, so confirm those three against your instrument before ordering.

  • Serving laboratories since 2022
  • Refurbished Instruments · Spare Parts Store · Service & AMC from the same team
  • GST-registered supplier · GSTIN 27ABIPZ1312R1Z0

Specifications

Nominal volume100 µL
BarrelGlass
PlungerStainless steel
Wearing partsPlunger seal and needle tip
Used onHPLC and GC autosamplers
Confirm before orderingNeedle length, gauge, tip style
Listed forAgilent, Shimadzu, Thermo Scientific, Waters